Philips Baby Monitor M2922A User Manual

M2922A, M2925A, M2926A  
Philips FM-2  
Portable Antepartum  
Fetal Monitor  
• 12-hours memory for fetal traces.  
Features  
• Internal battery power or AC power from external  
power supply.  
• 6-hours (approx.) battery back-up operation,  
rechargeable during normal monitoring.  
• Download capability to transmit traces stored in  
its memory to a PC or a Philips OB TraceVue  
surveillance and alerting system.  
• Online transmission to a monitoring system such  
as OB TraceVue.  
• Easily connects to off-the-shelf printer for  
documentation.  
Description  
FM-2 is a fetal monitor for non invasively  
measuring, and showing maternal uterine  
contractions and the fetal heart rate as a graphic  
display, and optionally on a strip chart recorder,  
printer or OB TraceVue system. The data is  
intended to aid in assessing the well being of the  
fetus during the final trimester of pregnancy non-  
stress test (NST).  
• Lightweight and portable, with carrying handle.  
• Easy to read display for viewing of FHR and Toco  
traces and numerics.  
• On-screen scrolling for viewing stored fetal traces.  
• 15 BPM separated grid lines for easy viewing of  
fetal heart rate accelerations for NST assessment.  
This portable fetal monitor offers a solution for  
antepartum screening in the private doctors office,  
outpatient clinics, and for remote applications such  
as traveling nurses visiting patients homes, or for  
clinics in rural areas seeking remote consultation.  
• Twins ultrasound monitoring capability.  
• Configurable audible and flashing fetal heart rate  
alarms.  
• Uses watertight transducers compatible with entire  
Series 50 range of fetal monitors.  
It is intended for use only by trained medical  
personnel located in hospitals, clinics, doctors  
offices and in the patients home or during transport  
under the direction of a licensed practitioner.  
• Event marker for easy documentation of events  
and kick counts.  
 
M1913A - Paper  
Monitoring Accessories  
Case of 40 packs thermal recorder paper, 0-100  
mmHg IUP scale, 50-210 FHR scale, 1 min  
(3cm) highlighted lines. Green. Not available in  
Europe.  
Standard Accessories and Supplies  
These are included with the monitor  
1 x Doppler Ultrasound FHR Transducer-  
1
M1356A  
Associated Products  
1 x External TOCO Transducer - M1355A  
1 x Remote Event Marker - 15249A  
External Recorder (M2925A)  
The recorder is an external device for printing stored  
fetal traces, or realtime fetal trace output. It connects  
to the FM-2 via a standard (IEEE 1284 compliant))  
printer cable.  
2 x Reusable Abdominal Belts - M1562A  
1 pack of three transducer knob adapters -  
M1356-43203  
FM-2 Viewer Utility Software (M2926A)  
1x 250ml ultrasound gel  
The monitor can send data to an IBM-compatible  
PC running Microsoft NT4 or 98.  
Users Guide  
Pocket Guide to Fetal Monitoring and Assessment  
by Susan Martin Tucker (English language  
shipments only). ISBN 0-323-00884-4,  
4th edition  
See the specifications section of this data sheet for  
system requirements.  
Mounting (M3904A and M3909A)  
M3904A - GCX Wallmounting.  
M3909A - GCX Rollstand  
Service Guide  
Additional Supplies and Accessories  
These are not included with the monitor, but can be  
purchased separately.  
M1356A - Second Doppler Ultrasound FHR  
Transducer for twins monitoring.  
M3919A - Canvas Tote Bag  
with room for monitor plus accessories.  
M1569A - Belt Fastening Button  
Pack of 10.  
M2208A - 100 Disposable Abdominal Belts  
1.3m long, 50mm wide (52long, 2wide). 1 box  
of 50 blue and 50 pink.  
40483A - Ultrasound Transmission Gel  
12 x 250 ml bottles, shelf life 24 months  
maximum, 6 months minimum.  
40483A - 5 Liter Refill Container  
shelf life 24 months maximum, 6 months  
minimum.  
M1910A - Paper  
Case of 40 packs thermal recorder paper, 0-100  
mmHg IUP scale, 30-240 FHR scale, 1 min  
(3cm) highlights. Orange (USA).  
M1911A - Paper  
Case of 40 packs thermal recorder paper, 0-100  
mmHg IUP scale, plus 0-13 kPa scale, 50-210  
FHR scale, no highlighted 3 cm lines. Green  
(International).  
1.The monitor has twins capability built in. To monitor  
twins you require an additional ultrasound transducer.  
Purchase additional M1356A.  
3
 
Recharge without  
monitoring:  
Maximum 8 hours to  
full recharge when no  
monitoring occurs (typical  
4.75 hours).  
Specifications  
Monitor (M2922A)  
Federal Law restricts this device to use by or on the  
Low power warning: Approximately 30 minutes  
1
at 25oC.  
order of a physician.  
Power - External Power Supply  
Standards Com pliance  
AC-powered  
110-120V~, 50-60Hz,  
0.15A  
or  
220-240V~, 50-60Hz,  
0.08A  
The FM-2 Portable Antepartum Fetal Monitor  
complies with the requirements of the Medical  
Device Directive 93/42/EEC and the following  
safety standards:  
EN60601-1/IEC60601-1, EN60601-1-1,  
EN60101-2-37  
Display  
IEC 1157  
IEC 60601-1  
UL2601-1  
CAN/CSA C22.2#601.1-M90  
Backlight:  
Active area:  
Cold cathode fluorescent  
3.78x 2.83”  
(320 x 240 dots)  
85 dots per inch  
10:1  
Resolution:  
Contrast:  
Class II equipment, double insulated  
Continuous operation  
Ultrasound  
Isata  
BPM range:  
Paper scale:  
:
< 4mW  
Input  
Connector  
Resulting  
Isolation  
Param eter  
USA paper: 30-240 BPM,  
International Paper: 50-210  
BPM.  
US  
Toco  
BF  
BF  
BF  
BF  
Ultrasound  
processing range: 50-240 BPM.  
Electrom agnetic Com patibility  
Accuracy:  
±1% ±1 BPM.  
Leakage current:  
10 µ A @ 264 VAC applied  
to transducer.  
Complies with EN60601-1-2  
Isolation:  
>4 kV RMS, Type BF  
applied part.  
Environm ental  
Operating Temp:  
Storage Temp:  
-10oC to 45oC  
Toco  
(50oF to 110oF)  
-20oC to 60oC  
(-29oF to 175oF)  
Range:  
0-100 relative units  
1 count  
±1% ±1 relative unit  
10 µ A @ 264 VAC applied  
to transducer  
>4 kV RMS, Type BF  
applied part  
Resolution:  
Accuracy:  
Leakage:  
Relative Humidity: 20% to 90% non  
condensing  
0-3048m (10,000 ft)  
Altitude:  
Isolation:  
Dim ensions  
Height:  
Width:  
Depth:  
Weight:  
15.24 cm (6.0 inches)  
26.4 cm (10.4 inches)  
13.7 cm (5.4 inches)  
2.27 kg (5lb)  
Transducers  
Doppler Ultrasound FHR (M1356A)  
Weight:  
185 g (6.5 oz)  
2.5m (8ft 2in)  
75mm (2.95in)  
Power - Battery  
Cable length:  
Diameter:  
Latex free  
Type:  
Operation:  
Lead acid, rechargeable.  
minimum 2-hours fully  
o
charged at 10 C, longer  
External Toco (M1355A)  
at higher temperature.  
Maximum 14 hours full  
recharge during monitor-  
ing (typical 11 hours).  
Transducer Weight: 180 g (6.3 oz)  
Cable length: 2.5m (8ft 2in)  
Diameter: 75mm (2.95in)  
Latex free  
Recharge during  
monitoring:  
Rem ote Event Marker (15249A)  
Length: 2.8m (9ft 2in)  
1.USA only  
4
 
Weight 75g (2.65oz)  
Latex free  
Width:  
Depth:  
Weight:  
23.4cm (9.2 inches)  
26.2 cm (10.3 inches)  
3.54 kg (7lb 13oz)  
Reusable Abdom inal Belt (M1562A)  
Standards Com pliance  
Length: 1.3m (52in)  
Width: 50mm (2in)  
Safety  
With button holes, and pre-attached buttons. Can  
be further cut. Contains latex.  
- Complies with EN60601-1, EN60601-1-1  
- Class I equipment, requires 3-prong outlet with  
protective earth  
FM-2 Viewer Utility Software  
(M2926A)  
- Continuous operation  
Electrom agnetic Com patibility  
Windows operating system-based utility software.  
Supplied on CD-ROM, together with operating  
instructions, license and 3m PC connection cable.  
Complies with EN60601-1-2  
Power  
System requirements:  
AC-powered  
110-120V~, 50-60Hz,  
0.15A  
or  
IBM-compatible PC (200MHz Pentium or  
equivalent)  
220-240V~, 50-60Hz,  
0.08A  
Standard VGA graphics adapter with 8MB  
memory  
Consumption:  
20 watts maximum at  
normal supply voltage  
CD-ROM drive  
60 MB (minimum) available space on hard drive  
Environm ental  
Windows 98, or NT (with service pack 6a)  
-10oC to 45oC  
(50oF to 110oF)  
-20oC to 60oC  
(-29oF to 175oF)  
operating system  
Operating Temp:  
Microsoft®, and Windows® are either registered  
trademarks or trademarks of Microsoft Corporation  
in the United States and/or other countries.  
Storage Temp:  
Relative Humidity: 20 to 90% non condensing  
Altitude: 0-3048m (10,000ft)  
The FM-2 viewer utility software complies with the  
requirements of the Medical Device Directive 93/  
42/EEC.  
Recorder Paper (Therm al)  
Federal Law restricts this device to use by or on the  
order of a physician.  
Style:  
Z-fold  
Pack size:  
End-of-pack:  
150 pages  
Mark along paper edge,  
stop sign on last page  
Front door, drop-in, self  
aligning  
Printers  
Supported printers are:  
HP DeskJet 990Cxi  
HP DeskJet 980Cxi  
HP DeskJet 960c  
HP DeskJet 840c  
HP DeskJet 640c  
HP DeskJet 350  
HP Photosmart 1215  
Loading:  
Paper detectors:  
Paper speeds:  
Paper out, and recorder  
loading door open  
Normal (1, 2, and 3 cm/  
min ±1%) and batch print  
(average 25 cm/min)  
±1% (exclusive of paper  
accuracy)  
Paper tracking  
accuracy:  
M1910A:  
30-240 FHR range, orange  
(Canada and USA) grid, no kPa scale, 3cm  
highlighted lines.  
This list will be updated periodically. Contact  
Philips Medical sales for details about currently  
supported printers.  
M1911A:  
(International)  
50-210 FHR range, green  
grid, kPa scale, no  
highlighted lines.  
M1913A/J:  
(Japan)  
50-210 FHR range, green  
grid, no kPa scale, 3 cm  
highlighted lines.  
Recorder (M2925A)  
Dim ensions  
Height:  
12.5 cm (4.9 inches)  
5
 
Philips offers a wide range of fetal monitors for  
different environments and applications. Review the  
capabilities of all fetal monitors to help you choose  
the monitor that is right for your needs.  
Feature  
Table legend:  
Standard  
o O ptional  
Unavailable  
Designed for portability/transport  
Integrated fetal trace display  
On screen NST assessment aid  
Battery operation and AC operation from power supply  
Off the shelf printer interface  
PC utility software  
External recorder  
Integrated recorder  
Fetal trace data storage (12 hours)  
o
Transmission to obstetrical surveillance system (for example  
OB TraceVue) via phone line  
Twins heart rate separation  
Direct system interface to obstetrical overview system such  
as OB TraceVue  
Watertight, easy to clean transducers  
Indication when two channels record the same heart rate  
(Cross Channel Verification - CCV)  
Indication when you accidentally record maternal SpO  
2
(Cross Channel Verification plus - CCV+)  
Performance PLUS! technology to ensure continuous,  
accurate and high-quality ultrasound traces  
Precision signal track and hold to reduce artifacts and  
increase sensitivity  
Interface to external fetal pulse oximeter (FSpO )  
o
o
o
o
o
2
Gross fetal body movement recording (fetal movement  
profile - FMP)  
External fetal measurements  
Single US FHR  
Twin US FHR  
Toco  
o
Internal fetal measurements  
DECG  
o
IUP/TTUIP  
Fetal SpO (FSpO )  
2
2
Integrated maternal  
measurements  
NIBP and SpO  
o
2
MECG  
ob  
Interface to external maternal monitor  
o
a.Series 50 XMO not available in USA or Japan  
b. When maternal parameters are not already integrated in the Series 50 XM/XMO monitor (that is, you have not ordered option A03)  
6
 
 
Philips Medical Systems is part of Royal Philips Electronics  
CO RPO RATE HEADQ UARTERS:  
Philips Medical Systems  
Nederland B.V.  
Postbus 10.000  
5680 DA Best  
Netherlands  
UNITED STATES:  
Philips Medical Systems  
Cardiac and Monitoring Systems  
3000 Minuteman Road  
Andover, MA 01810  
(800) 934-7372  
CANADA:  
Philips Medical Systems  
2660 Matheson Blvd. E.  
Mississauga, Ontario L4W 5M2  
(800) 291-6743  
EURO PE, MIDDLE EAST AND AFRICA:  
Philips Medizinsysteme Böblingen GmbH  
Cardiac and Monitoring Systems  
Hewlett-Packard Str. 2  
71034 Böblingen  
Germany  
Fax: (+49) 7031 463 1552  
LATIN AMERICA HEADQ UARTERS:  
Philips Medical Systems  
5200 Blue Lagoon Drive  
9th Floor  
Miami, FL 33126  
(305) 267-4220  
ASIA PACIFIC HEADQ UARTERS:  
Philips Medical Systems  
24F Cityplaza One  
1111 King's Road  
Taikoo Shing, Hong Kong  
(+852) 3197 7777  
© 2002 Philips Medizinsysteme Böblingen GmbH  
All rights are reserved. Reproduction in whole or in part is  
prohibited without the prior written consent of the copyright  
holder.  
Published April, 2002  
5990-0507EN  
 

Pentax Digital Camera K m K2000 User Manual
Philips Headphones HC8650 User Manual
Philips Model Vehicle 32PFL9606 User Manual
Philips Surge Protector SPP1188WA User Manual
Planar Computer Monitor SD2320W User Manual
Planet Technology Switch FSD 800 FSD 1600 FSD 801 FSD 1601SC FSD 1601 ST User Manual
Plantronics Headphones VOYAGER855 User Manual
Polycom CRT Television V700 User Manual
Polycom Telephone 32x User Manual
ProMaster Camera Accessories FL100 Canon User Manual